top of page

12 axis ultrasonic immersion system is designed for inspection of a complex structures and parts. The system is carefully designed to get high quality image.

 

Orbissys manufactures large scale multi-axis immersion scanning

systems for parts manufacturers and research institutes. 

The scanning dimension and water tank size are ordered

as customers request. 

Orbissys carefully design the frame structure and integrate electronics to get the best quality image resolution while minimizing noise from motion control circuitry.

 

● Feature:

- 12 motion axis(8 linear motion axis and 4 rotation motion axis)

- Precise sensor control for high frequency immersion test

- Water tank for immersion test

- Multi-channel Ultrasonic P/R

- Sensor control and data aqcuisition programing

- A, B, C-scan display and storage

- Acoustic beam profile measurement

 

● Composition of system:

- 12 axis motion stage

- Water tank

- Ultrasonic P/R

- Data aqcuisition board

- A, B, C-scan software

 

● 12 Axis motion stage:

X Axis:

Stroke: 1600mm

Repeatability: ≤100㎛

Min. Motion: ≤50㎛

Encoder Resolution: 10000 PPR

Y Axis:

Stroke: 1400mm

Repeatability: ≤100㎛

Min. Motion: ≤50㎛

Encoder Resolution: 10000 PPR

* This axis is actuated by two motors and ball screws synchronized

Z Axis:

Stroke: 1200mm

Repeatability: ≤100㎛

Min. Motion: ≤50㎛

Encoder Resolution: 10000 PPR

Rz Axis:

Motion angle: 360 degree

Repeatability: ≤5/100 degree

Min. Motion: ≤5/100 degree

Encoder Resolution: 10000 PPR

T Axis:

Motion angle: 60 degree

Repeatability: ≤5/100 degree

Min. Motion: ≤5/100 degree

Encoder Resolution: 10000 PPR

 

● Water tank:

Water volume: 2000mm(W) x 3000mm(L) x 1500mm(D)

Material: Stainless steel

Transparent window installed flange: 6

Level adjuster: more than 8

Water supply and sewage system: 1 each

* Water tank structure is sufficiently rigid to sustain high volume water.

 

● Control software:

Initial position programing and manual motion control

Synchronized control of motion, DAQ unit and pulse generator unit

Data acquisition and storage according to programed position table

Data acquisition and storage according to programed start/end position table

Sync signal output at programed position interval

Peak and TOF value measurement and storage

A-scan display windows are more than 8

C-scan display and storage form A-scan data

Acoustic beam profile measurement and storage 

 

 

 

bottom of page